Leon County Court March 6, 1991, Case No. 91-02069 (Consolidated) (State v. Numerous)
Also known as
United States · · · 06-03-1991
*90OPINION OF THE COURT
TERRY P. LEWIS, County Judge.
ORDER GRANTING IN PART DEFENDANTS’ MOTION IN LIMINE
THIS CAUSE came to be heard upon Motions in Limine to bar results of chemical breath tests. Upon the evidence presented, the Court makes the following findings:
THE FACTS
Each of these defendants, charged with D.U.I., submitted to a chemical breath test on a device known as the Intoximeter 3000, Serial Number 3707, Revision B-l (hereafter Intoximeter 3000). A component part of the instrument, known as the Taguchi Sensor (or T-Cell), measures acetone and other hydrocarbons so as to distinguish these substances from ethanol alcohol.
On January 11, 1990, prior to the test administered to the Defendants in these cases, the Taguchi Sensor in this instrument was “bypassed” by representatives of the Tallahassee Police Department and the Department of Health and Rehabilitative Services.1 The result of the bypass is that the instrument no longer measures or alerts the operator to the presence of acetone or other hydrocarbons.
ISSUE
IS THE BYPASS OF THE TAGUCHI SENSOR SUCH A
MODIFICATION OF THE INSTRUMENT SO THAT IT IS
NO LONGER AN “APPROVED” INSTRUMENT UNDER
FLORIDA STATUTES SECTION 316.1932(l)(b) AND
SECTION 10D-42.022, F.A.C.
Florida law requires a person who wishes to operate a motor vehicle in this State to submit to an approved chemical test in order to determine the alcohol content of his blood if there are reasonable grounds to believe he was operating a motor vehicle while under the influence of alcoholic beverages. Section 316.1932(l)(a), (Fla. Stat.). The Department of H.R.S. was delegated the responsibility of approving instruments and testing techniques. Accordingly, Rule 10D-42.022 F.A.C. requires prior approval of H.R.S. before an instrument can be *91used in this State. Rule 10D-42.023 F.A.C. contains an accuracy checking requirement prior to registration. Rule 10D-42.024(4-13) F.A.C. is a list of the instruments approved under Rule 10D-42.022 F.A.C.